引用本文:黄廷磊,李敏,郑刚,王乃宁,陈亚珠.一种改进的消光式颗粒计数传感器[J].中国激光,2004,31(6):49-751.
作者姓名:黄廷磊  李敏  郑刚  王乃宁  陈亚珠
作者单位:1. 桂林电子工业学院计算机系,广西,桂林,541004;上海交通大学生物医学仪器研究所,上海,200030
2. 桂林电子工业学院计算机系,广西,桂林,541004
3. 上海理工大学光电学院,上海,200093
4. 上海交通大学生物医学仪器研究所,上海,200030
摘    要:分析了消光式液体颗粒计数传感器的典型结构,并根据Mie理论计算了折射率为1.33的消光系数曲线,说明颗粒粒径在小于2μm时与消光系数呈多值性对应关系,指出这种多值性关系是影响此类计数器不能用于检测小颗粒的主要原因.最后提出一种改进的液体颗粒计数传感器结构,对传感器中光学部分的采光透镜组进行了改造,增加散射光探测器和光源监控探测器.通过计算说明了将散射光探测器安置在45°采光角时的原因,此时的光能与颗粒粒径之间有较好的单值对应关系,还可减少折射率对测量结果的影响.实验对比表明,改进结构的传感器对小于2μm的标准颗粒试样有较好的检测能力,而且多次重复测量结果相近;对大于2μm颗粒的检测能力仍然保持着较高的水平.

关 键 词:激光技术  消光  光学颗粒计数器  光散射
收稿时间:2003/3/31

An Improved Sensor of Extinction Particle Counter
HUANG Ting-lei ,LI Min,ZHENG Gang,WANG Nai-ning,CHEN Ya-zhu.An Improved Sensor of Extinction Particle Counter[J].Chinese Journal of Lasers,2004,31(6):49-751.
Authors:HUANG Ting-lei    LI Min  ZHENG Gang  WANG Nai-ning  CHEN Ya-zhu
Affiliation:HUANG Ting-lei 1,2,LI Min1,ZHENG Gang3,WANG Nai-ning3,CHEN Ya-zhu2 1Department of Computer,Guilin University of Electronic Technology,Guilin,Guangxi 541004,China 2College of Life Science and Technology,Shanghai Jiaotong University,Shanghai 200030,China 3College of Opto-Electronics,University of Shanghai for Science and Technology,Shanghai 200093,China
Abstract:Particle counter of extinction type is used mainly to detect the particulate contamination of larger sizes, and widely applied in contamination control measurement and monitor, whereas it is unable to detect smaller sizes particulate, the reason is discussed in this paper. Based on Mie theory, extinction coefficient is calculated, at the same time, the representative structure of this type sensor is analyzed. Numerous calculation and analysis showed that it is due to the ambiguity relationship of extinction coefficient versus particulate sizes less than 2 μm and unreasonable design. An improved design of the sensor was proposed, which employs a specially designed detecting lens unit to collect more scattered light and a monitoring device to control the light source, the lens of collecting scattering light is arranged at 45 degree, which can lead the good monodromy for light energy versus particulate sizes and less interference due to particle refractive index. Comparative experiments showed that the lower limit of the measuring size range could be extended by the new sensor design. The new sensor has a good detect ability for particle sizes less than 2 μm and lower repetition errors, and furthermore, it still has a high level for detecting larger sizes particle, it suits for more application of contamination degree monitoring.
Keywords:laser technique  extinction light  extinction particle counter (EPC)  light scattering