应用OBI系统分析胸部肿瘤调强放疗的摆位误差
引用本文:蒋琳,陈晓品,金丹,宋延波,刘丽. 应用OBI系统分析胸部肿瘤调强放疗的摆位误差[J]. 激光杂志, 2009, 30(5): 92-93
作者姓名:蒋琳  陈晓品  金丹  宋延波  刘丽
作者单位:重庆医科大学附属第一医院肿瘤科,重庆,400016;重庆医科大学附属第一医院肿瘤科,重庆,400016;重庆医科大学附属第一医院肿瘤科,重庆,400016;重庆医科大学附属第一医院肿瘤科,重庆,400016;重庆医科大学附属第一医院肿瘤科,重庆,400016
摘    要:目的:应用机载影像系统(on boardinmger,OBl)分析胸部肿瘤调强放射治疗时分次间的摆位误差,并依此计算CTV外扩PTV边界的大小。方法:应用Varian-21EX医用直线加速器治疗47例胸部肿瘤。每周治疗前获取锥形束断层扫描(cone beam computed tomography,CBCT)图像,将CBCT图像和计划CT图像及其靶中心匹配,计算平移和旋转误差。结果:平移误差左右(x)方向为(-0.3194±3.6943)mm,头脚(y)方向为(0.4851±6.2636)mm,前后(z)方向为(0.4144±5.2818)mm;旋转误差冠状面(c)为(0.2080±1.4925)°,矢状面(s)为(0.0922±1.2976)°结沦:对于胸部肿瘤调强放疗(IMRT),临床靶体积(CTV)到计划靶体积(PTV)的外放边界在左右方向宜为7mm,头脚方向宦为11mm,前后方向宜为10mm。考虑到旋转误差,与靶区比较长时靶区两端外放要更大一些.

关 键 词:摆位误差  OBI系统  CBCT  调强放射治疗

An analysis of setup error with on board imager for chest tumors treated with intensity modulated radiation therapy
JIANG Lin,CHEN Xiao - pin,JIN Dan,SONG Yan-bo,LIU Li. An analysis of setup error with on board imager for chest tumors treated with intensity modulated radiation therapy[J]. Laser Journal, 2009, 30(5): 92-93
Authors:JIANG Lin  CHEN Xiao - pin  JIN Dan  SONG Yan-bo  LIU Li
Affiliation:( Department of Radiation Oncology, the First Affiliated Hospital, Chongqing Medical University, Chongqing 400016, China)
Abstract:Objective:To analyze the setup errors by on board imager in chest tumor treated with Intensity Modulated Radiotherapy for calculating safety margins of FTV refer to CTV. Metheds:47 patients with chest tumor were treated with Varian - 21EX Linear accelerator. All patients had received cone beam computed tomography(CBCT)before radiotherapy was delivered once a week.The acquired images of cone heron computed tomography were co - registered with planning CT and errors of isocentcr positions on x,y,z axes and rotational deviations were obtained and analyzed. Results:The translational deviations were( - 0.3194 ± 3.6943)mm, (0.4851 ± 6.2636)mm and(0.4144 ± 5.2818)mm in medi - lateral, cranio - caudal and anterior - posteriar directions, respectively. The rotational deviations were(0.2080± 1.4925)°and(0.0922 ±1.2976)°in sagittal plane and coronal plane, respectively. Conclusions: For the patients with chest tumors undergoing intensity modulated radiation therapy, the margins between the CTV and PTV should be 7 mm in lateral direction, 11 mm in cranio- caudal direction and 10 mm in anterior - posterior direction. Considering thc rotational variations, the margin should be expanded at the end of the target.
Keywords:CBCT