Incoherent-scatter computed tomography with monochromaticsynchrotron x ray: feasibility of multi-CT imaging system forsimultaneous measurement-of fluorescent and incoherent scatter x rays
Authors:Yuasa  T Akiba  M Takeda  T Kazama  M Hoshino  A Watanabe  Y Hyodo  K Dilmanian  FA Akatsuka  T Itai  Y
Affiliation:Fac. of Eng., Yamagata Univ., Yonezawa;
Abstract:We describe a new system of incoherent scatter computed tomography (ISCT) using monochromatic synchrotron X rays, and we discuss its potential to be used in in vivo imaging for medical use. The system operates on the basis of computed tomography (CT) of the first generation. The reconstruction method for ISCT uses the least squares method with singular value decomposition. The research was carried out at the BLNE-5A bending magnet beam line of the Tristan Accumulation Ring in KEK, Japan. An acrylic cylindrical phantom of 20-mm diameter containing a cross-shaped channel was imaged. The channel was filled with a diluted iodine solution with a concentration of 200 μgI/ml. Spectra obtained with the system's high purity germanium (HPGe) detector separated the incoherent X-ray line from the other notable peaks, i.e., the iKα and Kβ1 X-ray fluorescent lines and the coherent scattering peak. CT images were reconstructed from projections generated by integrating the counts In the energy window centering around the incoherent scattering peak and whose width was approximately 2 keV. The reconstruction routine employed an X-ray attenuation correction algorithm. The resulting image showed more homogeneity than one without the attenuation correction