医用级硅橡胶残存单体及低聚物的测定
引用本文:刘伟,王传栋,郑兴仪. 医用级硅橡胶残存单体及低聚物的测定[J]. 有机硅材料, 1999, 13(1): 18-19
作者姓名:刘伟  王传栋  郑兴仪
作者单位:山东省医疗器械研究所,济南,250100
摘    要:选择Si-C伸展振动和CH3平面摇摆震动在~855cm-1及~800cm-1处的吸收峰作为分析峰,采用红外光谱分析法对SYZ—2医用级硅橡胶残存单体及低聚物含量作了测定,未检出残存单体及低聚物。

关 键 词:单体  低聚物  医用硅橡胶  红外光谱

Determination of the Residue of the Monomer and Oligomer in Medical Silicone Rubber
Liu Wei,Wang Chuandong,Zheng Xingyi. Determination of the Residue of the Monomer and Oligomer in Medical Silicone Rubber[J]. Silicone Material, 1999, 13(1): 18-19
Authors:Liu Wei  Wang Chuandong  Zheng Xingyi
Abstract:The silicone rubber SYZ - 2 was used as the assistant material of the female long - acting contraceptive subdermal implant as well as plastic saddle nose. The quality of SYZ - 2 must be controlled strictly because it is a biomaterial covered in human body for long time. Selected absorption peak at 855 cm -1 and 800 cm -1 as analysis peak, the residues of the monomer and oligomer in SYZ - 2 were detected by IR analysis.
Keywords:silicone rubber monomer oligomer IR analysis