首页 | 本学科首页   官方微博 | 高级检索  
     

石墨烯修饰电极同时测定邻苯二酚和对苯二酚
引用本文:万其进,廖华玲,刘义,魏薇,舒好,杨年俊.石墨烯修饰电极同时测定邻苯二酚和对苯二酚[J].武汉工程大学学报,2013,35(2):16-23.
作者姓名:万其进  廖华玲  刘义  魏薇  舒好  杨年俊
作者单位:武汉工程大学绿色化工过程教育部重点实验室,湖北省新型反应器与绿色化学工艺重点实验室,湖北武汉430074
基金项目:国家自然科学基金(21075096,21275113)
摘    要:制备石墨烯玻碳修饰电极,进而采用循环伏安法、交流阻抗等电化学方法对该电极进行表征,研究该石墨烯修饰电极在邻苯二酚和对苯二酚上的电化学行为.结果表明,在石墨烯修饰电极上邻苯二酚的氧化峰电位和还原峰电位分别是270mV和161mV,对苯二酚氧化峰电位和还原峰电位分别是145mV和64mV,由于邻苯二酚和对苯二酚的氧化峰电位大约相离125mV,还原峰大约相离97mV,因此适合同时检测邻苯二酚和对苯二酚.邻苯二酚和对苯二酚的浓度在5.0×10-6~1.0×10-4mol/L范围内与峰电流分别呈良好的线性关系;且在8.0×10-5~1.0×10-3mol/L范围能同时检测邻苯二酚和对苯二酚,邻苯二酚的检测限可达5.0×10~7mol/L,对苯二酚的检测限可达1.0×10-mol/L.该石墨烯修饰电极可作为电化学传感器用于邻苯二酚和对苯二酚的含量同时测定及环境水体中实际样品的分析.

关 键 词:石墨烯  修饰电极  示差脉冲法  邻苯二酚  对苯二酚

Simultaneous determination of catechol and hydroquinone in graphene modified electrode
Authors:WAN Qi-jin  LIAO Hua-ling  LIU Yi  WEI Wei  SHU Hao  YANG Nian-jun
Affiliation:(School of chemical Engineering and Pharmacy,Wuhan Institude of Technology, Key Lab for Green Chemical Process of Ministry of Education,Wuhan 430074,China)
Abstract:A novel graphene modified glassy carbon electrode was fabricated. The resulting substrates were characterized by Cyclic Voltammetry and EIS in -Fe (CN)6]3-/4- solution and showed the electrochemical behavior of catechol and hydroquinone on the graphene modified glassy carbon electrode. Experiment result shows that the catechol oxidation peak potential is 270 mV and reduction peak potential is 161 mV, and the hydroquinone oxidation peak potential is 145 mV and reduction peak potential is 64 mV on the graphene modified electrode, respectively. The oxidation peak potential distance is about 125 mV and the reduction peak potential distance is about 97 mV of catechol and hydroquinone which are suited for the simultaneous detection. Catechol and hydroquinone have good electrocatalytic activity on modified electrode and the peak currents of differential pulse voltammetry are liner to the catechol and hydroquinone over the range of 5.0 ×10-8 -1.0× 10-4 mol/L, respectively, and the graphene modified electrode can simultaneously detect catechol and hydroquinone in the rang of 8.0× 10-5-1. 0 × 10-3 mol/L. The catechol detection limit is 5.0 × 10-7 mol/L, the hydroquinone detection limit is 1.0 × 10-7 mol/L. So the graphene modified electrode can be used for analysis the facilitation of actual samples and electrochemical sensors and biosensors.
Keywords:graphenes  modified electrodes  differential pulse methods catechols  hydroquinones  
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号