Detection and validation of the body edge in low count emission tomography images
Affiliation:1. Los Angeles Biomedical Research Institute, 1124 W Carson Street, Torrance 90502, California;2. University of Michigan, Ann Arbor, Michigan;3. Long Beach Memorial, Long Beach, California;4. Hofstra North Shore – LIJ School of Medicine, Hempstead, New York;5. Emory University, Atlanta, Georgia
Abstract:Segmentation of the body edge in tomographic images with low count, noisy edges is needed for both PET and SPECT respiratory motion correction and brain SPECT attenuation correction. To reduce noise we re-projected tomographic images and searched for edges in the projection count profiles or their spatial derivatives. Edge location versus projection angle was fitted with cosine basis functions after rejecting outliers and including information about edges of previous sections. We processed 10 s duration FDG PET of the thorax, HMPAO brain, DAT brain and lung perfusion SPECT. Stable edges for all four types of scan were achieved but with different profiles. Edges were validated against edges of coregistered CT or MRI. The best root mean square (rms) accuracy was 8.2 mm (PET) and 5.2 mm (brain SPECT). Inter-scan variability (standard deviation) in the estimated-to-control edge distance for 17 PET scans was 0.4 mm, for 25 ordered subset expectation maximisation (OSEM) reconstructed brain SPECT 1.0 mm and for 18 filtered back-projection (FBP) reconstructed brain SPECT 1.4 mm.