首页 | 本学科首页   官方微博 | 高级检索  
     


Temperature dependence of retention in reversed-phase liquid chromatography. 1. Stationary-phase considerations.
Authors:L A Cole  J G Dorsey
Affiliation:Department of Chemistry, University of Cincinnati, Ohio 45221-0172.
Abstract:The retention mechanism in reversed-phase liquid chromatography (RPLC) has been investigated by examining the temperature dependence of retention, with emphasis on the role of the stationary phase in the retention process. Both chromatographic temperature studies and differential scanning calorimetry were used to examine the role of alkyl chain bonding density on the retention mechanism in RPLC. Phase transitions of reversed-phase stationary phases were observed at bonding densities greater than 2.84 mumol/m2. Thermodynamic constants for the transfer of a solute from the mobile phase to the stationary phase (delta H degrees and delta S degrees) were calculated for low bonding density columns, and comparison of these values to previously reported values for the partitioning of a nonpolar solute from the bulk organic liquid to water indicated that the chromatographic retention process is not well-modeled by bulk-phase oil-water partitioning processes. In addition, this data showed that the entropic contribution to retention becomes more significant with respect to the enthalpic contribution as the stationary-phase bonding density is increased, providing additional support that partitioning, rather than adsorption, is the relevant model of retention.
Keywords: