首页 | 本学科首页   官方微博 | 高级检索  
     


A low-noise current-sensitive preamplifier for X-ray computed tomography with applying a charge-sensitive preamplifier
Authors:Ikuo Kanno  Hideaki Onabe
Affiliation:1. Department of Nuclear Engineering, Graduate School of Engineering, Kyoto University , Kyoto, Japan kanno@nucleng.kyoto-u.ac.jp;3. Raytech Corporation , Utsunomiya, Tochigi, Japan
Abstract:ABSTRACT

A low-noise current-sensitive preamplifier was developed for aiming low-dose X-ray computed tomography with applying a charge-sensitive preamplifier. We compared the number of X-ray photons needed for current measurement with the new preamplifier with that needed with a conventional preamplifier. Also, the dose rates for above conditions were measured by a commercial dose rate meter. Results show that the new preamplifier reduced the number of X-ray photons needed by two orders of magnitude compared with the conventional one.
Keywords:X-ray  current measurement  low-noise preamplifier  computed tomography  dose rate  experiment