Na188ReO4溶液活度测量的标准化研究
引用本文:钟建国,弓全胜,黄清泉,姜金岭.Na188ReO4溶液活度测量的标准化研究[J].同位素,2008,21(3):173-177.
作者姓名:钟建国  弓全胜  黄清泉  姜金岭
作者单位:中国药品生物制品检定所,北京,100050
基金项目:科技部发展司科兴贸项目
摘    要:摘要:治疗用放射性药品活度与其用药安全性和有效直接相关。为了确保含188Re放射性药品用药安全有效,本工作利用Na188ReO4溶液,对使用活度计测量Na188ReO4 溶液活度的方法进行了标准化研究。利用4π液闪、4πβ-γ符合方法,通过三家单位4套装置测量,确定Na188ReO4溶液标准源的活度量值;再利用已知活度的标准源,对使用活度计测量Na188ReO4 溶液活度的方法进行了标准化研究,确定了测量条件;最后在三个地区的10台不同活度计上进行了活度测量比对试验,结果表明,利用CRC系列活度计测量Na188ReO4溶液活度,各测定值均在标准值的±3%以内。因此,进行标准化标定后,活度计适于测量Na188ReO4溶液的活度。

关 键 词:Na188ReO4  溶液  活度测量  标准化
收稿时间:2008-03-18

Study on the Standardization for the Radioactivity Measurement for Sodium Perrhenate 188Re Solution
ZHONG Jian-guo,GONG Quan-sheng,HUANG Qing-quan,JIANG Jin-ling.Study on the Standardization for the Radioactivity Measurement for Sodium Perrhenate 188Re Solution[J].Isotopes,2008,21(3):173-177.
Authors:ZHONG Jian-guo  GONG Quan-sheng  HUANG Qing-quan  JIANG Jin-ling
Abstract:Abstract:The safety and efficacy of the therapeutic radiopharmaceuticals directly depends on its dosage. In order to determine the radioactivity of 188Re radiopharmaceuticals accurately with dose calibrators , and ensure the safety and efficacy of the 188Re radiopharmaceuticals, the standardization for the radioactivity measurement of Sodium Perrhenate 188Re Solution with dose calibrators has been studied, in which 4π liquid scintillator and 4πβ-γ coincidence method are used to determine the values of radioactivity for the Sodium Perrhenate 188Re Solution standard source with four experimental set-ups in three laboratories, then the dose calibrators are calibrated with the standard source, and the effect of the type of dose calibrator, the container of the standard source, and the volume of the solution are studied systematically, at last the comparative test is carried out among 10 different dose calibrators in three cities. The results show that the values of CRC series dose calibrators are all in coincidence within ±3%.In conclusion, having been calibrated by standardization, the CRC series dose calibrators can be used to accurately determine the radioactivity of Sodium Perrhenate 188Re Solution.
Keywords:Key words: Na188ReO4 solution  radioactivity measurement  standardization