Experimental and theoretical comparison of NIR spectroscopy measurements of cerebral hemoglobin changes
Authors:M Firbank  CE Elwell  CE Cooper  DT Delpy
Affiliation:Department of Medical Physics and Bioengineering, University College London, London WC1E 6JA, United Kingdom CO4 3SQ.
Abstract:Two near-infrared spectroscopy (NIRS) methods are available for measuring changes (Delta) in total cerebral hemoglobin concentration (CHC): 1) a continuous measurement of the changes in total hemoglobin concentration (Delta[Hb]tot) and 2) the difference between two absolute measurements of CHC, each derived from a small, controlled change in inspired O2 fraction. This paper investigates the internal consistency of these two methods by using an experimental and theoretical comparison. NIRS was used to measure [Hb]tot in five newborn piglets before and after a change in arterial PCO2. Delta[Hb]tot demonstrated a low coefficient of variation of 2.8 +/- 2.8 (SD) % which allowed changes in CO2-cerebral blood volume reactivity to be clearly discriminated. However, a high coefficient of variation of 22.8 +/- 3.5% on the DeltaCHC measurements obscured any CO2 reactivity changes. A theoretical analysis demonstrates the effects of optical pathlength, background absorption, scatter, and blood vessel diameter on both methods. For more accurate monitoring of CHC, individual measurements of optical pathlength and more accurate pulse oximetry are required.