Optimization of a dual-rotating-retarder Mueller matrix polarimeter |
| |
Authors: | Smith Matthew H |
| |
Affiliation: | JDS Uniphase Corporation, San Jose, California 95131, USA. matt.smith@jdsu.com |
| |
Abstract: | ![]() The dual-rotating-retarder configuration is one of the most common forms of the Mueller matrix polarimeter. I perform an optimization of this polarimeter configuration by minimizing the condition number of the system data reduction matrix. I find the optimum retardance for the rotating retarders to be 127 degrees. If exactly 16 intensity measurements are used for a Mueller matrix calculation, a complex relationship exists between the condition number and the sizes of the angular increments of the two retarders. If many intensity measurements are made, thus overspecifying the calculation, I find broad optimal ranges of angular increments of the two retarders that yield essentially equal performance. Experimental results are given. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|