首页 | 本学科首页   官方微博 | 高级检索  
     


Error analysis of the parameters for non-coaxial grazing X-ray microscope
Authors:Zhao Lingling  Hu Jiasheng and Li Xiang
Affiliation:(1) School of Electronic and Information Engineering, Dalian University of Technology, Dalian, 116024, China;(2) School of Physics and Electronic Engineering, Ludong University, Yantai, 264025, China
Abstract:A non-coaxial grazing X-ray microscope, consisting of four spherical mirrors, is designed for diagnosis of inertial confinement fusion (ICF). The aberrations and imaging quality of the microscope are analyzed. To acquire excellent imaging quality, suitable tolerances for manufacturing and assembling the microscope are necessary. This paper researches the changes of Gauss parameters and aberrations due to component and subsystem parameters (such as the radius of the mirror, angle between mirrors, grazing angle, object distance, etc.). Here, spot diagrams and modulation transfer function (MTF) are first adopted to quantitatively evaluate the imaging quality of the microscope. Suitable manufacturing tolerances of components and assembly tolerances of the system are established on the basis of the discussion and analysis. A set of non-coaxial grazing X-ray microscopes is manufactured based on the tolerances. In site tests of ICF, the images with high resolution are obtained by the microscope. __________ Translated from Optics and Precision Engineering, 2006, 14(1): 34–42 译自: 光学精密工程]
Keywords:X-ray imaging  X-ray microscope  tolerance  error analysis
本文献已被 万方数据 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号