High precision axial coordinate readout for an axial 3-D PET detector module using a wave length shifter strip matrix
Authors:A Braem  E Chesi  C Joram  J Sguinot  P Weilhammer  R De Leo  E Nappi  W Lustermann  D Schinzel  I Johnson  D Renker  S Albrecht
Affiliation:

aCERN, PH Department, CH-1211 Geneva, Switzerland

bINFN, Sezione di Bari, I-70122 Bari, Italy

cETH Zürich, CH-8092 Zürich, Switzerland

dPaul Scherrer Institut, CH-5232 Villigen, Switzerland

eUniversity Hospital Geneva, CH-1211 Geneva, Switzerland

Abstract:We describe a novel method to extract the axial coordinate from a matrix of long axially oriented crystals, which is based on wavelength shifting (WLS) plastic strips. The method allows building compact 3-D axial gamma detector modules for PET scanners with excellent 3-D spatial, timing and energy resolution while keeping the number of readout channels reasonably low. A voxel resolution of about 10 mm3 is expected. We assess the performance of the method in two independent ways, using classical PMTs and G-APDs to read out the LYSO (LSO) scintillation crystals and the WLS strips. We observe yields in excess of 35 photoelectrons from the strips for a 511 keV gamma and reconstruct the axial coordinate with a precision of about 2.5 mm (FWHM).
Keywords:PET  Santillator  HPD  G-APD