High precision axial coordinate readout for an axial 3-D PET detector module using a wave length shifter strip matrix
Authors:A. Braem, E. Chesi, C. Joram, J. S  guinot, P. Weilhammer, R. De Leo, E. Nappi, W. Lustermann, D. Schinzel, I. Johnson, D. Renker,S. Albrecht
Affiliation:

aCERN, PH Department, CH-1211 Geneva, Switzerland

bINFN, Sezione di Bari, I-70122 Bari, Italy

cETH Zürich, CH-8092 Zürich, Switzerland

dPaul Scherrer Institut, CH-5232 Villigen, Switzerland

eUniversity Hospital Geneva, CH-1211 Geneva, Switzerland

Abstract:We describe a novel method to extract the axial coordinate from a matrix of long axially oriented crystals, which is based on wavelength shifting (WLS) plastic strips. The method allows building compact 3-D axial gamma detector modules for PET scanners with excellent 3-D spatial, timing and energy resolution while keeping the number of readout channels reasonably low. A voxel resolution of about 10 mm3 is expected. We assess the performance of the method in two independent ways, using classical PMTs and G-APDs to read out the LYSO (LSO) scintillation crystals and the WLS strips. We observe yields in excess of 35 photoelectrons from the strips for a 511 keV gamma and reconstruct the axial coordinate with a precision of about 2.5 mm (FWHM).
Keywords:PET   Santillator   HPD   G-APD