首页 | 本学科首页   官方微博 | 高级检索  
     


Optical microcantilever consisting of channel waveguide for scanning near-field optical microscopy controlled by atomic force
Authors:T Niwa  Y Mitsuoka  K Kato  S Ichihara  N Chiba  M Shin-Ogi  K Nakajima  H Muramatsu  & T Sakuhara
Affiliation:Corporate R &D, Seiko Instruments Inc., 563 Takatsuka-Shinden, Matsudo-shi, Chiba 270-2222, Japan
Abstract:We develop a novel optical microcantilever for scanning near-field optical microscopy controlled by atomic force mode (SNOM/AFM). The optical microcantilever has the bent channel waveguide, the corner of which acts as aperture with a large tip angle. The resonance frequency of the optical microcantilever is 9 kHz, and the spring constant is estimated to be 0.59 N/m. The optical microcantilever can be operated in contact mode of SNOM/AFM and we obtain the optical resolution of about 200 nm, which is as same size as the diameter of aperture. We confirm that the throughput of optical microcantilever with an aperture of 170 nm diameter would be improved to be more than 10?5.
Keywords:Atomic force microscopy  optical microcantilever  scanning near-field optical microscopy  waveguide  
设为首页 | 免责声明 | 关于勤云 | 加入收藏