A method to modify coordinates of detectors in positron emission tomography systems
Authors:H. Sabet   K. Ishii   S. Matsuyama   Y. Kikuchi   K. Nakazawa   A. Esmaili Torshabi  H. Yamazaki
Affiliation:aDepartment of Quantum Science and Energy Engineering, Graduate School of Engineering, Tohoku University, Aoba-ku, Aramaki Aza-Aoba 6-6-1, Sendai 980-8579, Japan;bCyclotron and Radioisotope Center, Tohoku University, Aoba-ku, Aramaki Aza-Aoba 6-3, Sendai 980-8579, Japan
Abstract:A new method to modify coordinates of detectors in any positron emission tomography (PET) system using a radioactive point source is developed. This method is based on selecting a centered detector in each detector block of PET and measuring coincidence counts between the two centered detectors in opposite detector blocks to find the coordinates of their LOR (Line of Response). Due to slight misalignment of detector positions, measured LORs may not intersect at a single point. Based on the proposed method, the coordinates of detectors can be measured with very high accuracy and the coordinate of the center of the gantry (which is normally the same as the center of field of view) can be defined correctly. The results of the application of our method to a small animal PET system (FinePET), which was recently developed at Tohoku University, Japan, are shown here. The method is expected to contribute to the design and development of PET systems which can realize a very high spatial resolution of less than 1 mm FWHM.
Keywords:Positron emission tomograph   Spatial resolution   Detector coordinate