首页 | 本学科首页   官方微博 | 高级检索  
     


Use of a low refractive index prism in surface plasmon resonance biosensing
Authors:Gaurav  Mitsunori  Yoshikazu  Jun  
Affiliation:

aGraduate School of Electronic Science and Technology, Shizuoka University, 3-5-1 Johoku, Hamamatsu, Shizuoka 432-8561, Japan

bResearch Institute of Electronics, Shizuoka University, 3-5-1 Johoku, Hamamatsu, Shizuoka 432-8561, Japan

cFaculty of Engineering, Shizuoka University, 3-5-1 Johoku, Hamamatsu, Shizuoka 432-8561, Japan

dGraduate School of Science and Technology, Shizuoka University, 3-5-1 Johoku, Hamamatsu, Shizuoka 432-8561, Japan

Abstract:A way to improve the angular sensitivity of surface plasmon resonance (SPR) biosensor by tuning the resonance position to a higher incident angle region using a lower refractive index glass prism is described in this paper. A novel effective 3-layer (E3L) model is described to transform a multiple layer biosensor configuration in context of SPR condition. The E3L model supports the use of a low refractive index prism for biosensing. The performance of the sensor in immunosensing is checked for two glass prisms of different refractive index materials. The experimental results showed an enhancement in the amount of resonance angle shift of the immunosensor for the lower refractive index glass prism.
Keywords:Surface plasmon resonance  Biosensor  Refractive index  Glass prism  Effective 3-layer model