首页 | 本学科首页   官方微博 | 高级检索  
     

DeterminationofNi,Ca,Mg,CuandPbinWolframbyInductivelyCoupledPlasmaAtomicEmisionSpectrometry
引用本文:王春梅,张淑珍. DeterminationofNi,Ca,Mg,CuandPbinWolframbyInductivelyCoupledPlasmaAtomicEmisionSpectrometry[J]. 稀有金属(英文版), 1997, 0(1)
作者姓名:王春梅  张淑珍
作者单位:Wang Chunmei and Zhang ShuzhenThe 46th Institute of Electronic Industry Ministry,Tianjin 300192,China
摘    要:DeterminationofNi,Ca,Mg,CuandPbinWolframbyInductivelyCoupledPlasmaAtomicEmisionSpectrometryWangChunmeiandZhangShuzhen(王春梅)(张淑...


Determination of Ni, Ca, Mg, Cu and Pb in Wolfram byInductively Coupled Plasma Atomic Emission Spectrometry
Wang Chunmei and Zhang ShuzhenThe th Institute of Electronic Industry Ministry,Tianjin ,China. Determination of Ni, Ca, Mg, Cu and Pb in Wolfram byInductively Coupled Plasma Atomic Emission Spectrometry[J]. Rare Metals, 1997, 0(1)
Authors:Wang Chunmei  Zhang ShuzhenThe th Institute of Electronic Industry Ministry  Tianjin   China
Affiliation:Wang Chunmei and Zhang ShuzhenThe 46th Institute of Electronic Industry Ministry,Tianjin 300192,China
Abstract:The method for the determination of Ni, Ca, Mg, Cu and Pb in wolfram wire by ICP AES was studied. Results showed that perhydrol affects the intensities of elements (Ni, Ca, Mg, Cu, Pb) and this effect can be compensated by internal reference element. It is also showed that the stability of plasma caused by perhydrol matrix was effectively improved. The replicate experiment and recovery experiment revealed perfect precision and accuracy. Results got by two methods were just the same.
Keywords:Wolfram   Trace elements   Atomic emission spectrometry
本文献已被 CNKI 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号