首页 | 本学科首页   官方微博 | 高级检索  
     


MR-guided focused-ultrasound ablation system: determination of precision and accuracy in preparation for clinical trials
Authors:KR Kendell  T Wu  JP Felmlee  BD Lewis  RL Ehman
Affiliation:Department of Radiology, MR Research Lab, Mayo Clinic, Rochester, MN 55905, USA.
Abstract:The authors analyzed the accuracy and precision of focal-spot placement with a magnetic resonance-guided, focused-ultrasound system. Average absolute accuracy errors ranged from 0.2 to 1.0 mm, and average absolute individual precision errors ranged from 0.2 to 0.3 mm. To prevent damage to vital structures, single sonications and sonication grids should be placed beyond approximately 2 and 1 mm, respectively.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号