首页 | 本学科首页   官方微博 | 高级检索  
     


Rainbow refractometry: simultaneous measurement of temperature, refractive index, and size of droplets
Authors:Hom James  Chigier Norman
Affiliation:Department of Mechanical Engineering, Carnegie Mellon University, Pittsburgh, Pennsylvania 15213-3800, USA. drjh4x@yahoo.com
Abstract:The capabilities and limitations of rainbow refractometry, specifically the minimum measurable droplet size and the errors in the refractive index (temperature), have been studied. We evaluate what we believe is a new method of indirectly applying the Lorenz-Mie theory to rainbow refractometry. The results show that this new method reduces the errors and eliminates the biases that may occur if the Airy theory is used. A more precise method to filter the high-frequency oscillations associated with the measurement signals was developed. Finally, it was discovered that the errors associated with rainbow refractometry are such that a single droplet measurement is unreliable. A mean refractive index should be determined on the basis of multiple droplet measurements.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号