The Impact of Isotopic Concentration,Impurities, and Cell Aging on the Water Triple-Point Temperature |
| |
Authors: | G F Strouse M Zhao |
| |
Affiliation: | (1) National Institute of Standards and Technology, 100 Bureau Dr., MS 8363, Gaithersburg, MD 20899, USA;(2) Hart Scientific Division, Fluke Corporation, American Fork, UT 84003, USA |
| |
Abstract: | In 2005, the National Institutes of Standards and Technology (NIST) and Fluke’s Hart Scientific Division initiated a study
to validate the isotopic correction algorithm applied to the realization temperature of triple point of water (TPW) cells.
Additionally, the study quantified the impact of water sample impurities on the TPW cell realization temperature. For this
study, eight TPW cells containing water of the same nominal isotopic concentration as Vienna Standard Mean Ocean Water (VSMOW)
were used. Five of the cells were manufactured with fused-quartz envelopes and the remaining three with borosilicate envelopes.
One TPW cell of each type was uniquely designed so that water samples could be periodically removed to analyze the isotopic
composition and to monitor any changes in water purity with time and thereby correlate changes in composition with changes
in realization temperature. The borosilicate TPW cells gave an average drift of −13 μK · yr−1 and the more stable fused-quartz TPW cells gave an average drift of −2 μK · yr−1. |
| |
Keywords: | Isotopic concentration ITS-90 TPW (triple point of water) cell Vienna standard mean ocean water (VSMOW) Water impurities Water triple point |