Eliminating the main source of instability and turbulence in TSI's 3071 DMA
Authors:Juan Fernandez de la Mora  Patrick Gao  Luis J. Perez-Lorenzo
Affiliation:Mechanical Engineering Department, Yale University, New Haven, Connecticut, USA
Abstract:The transfer function of TSI's widely used 3071 Differential Mobility Analyzer (DMA) widens drastically when the flow rate Q of sheath gas exceeds 30–40 lit/min, limiting its ability to resolve very small particles. This flow instability is unexpected at the prevailing relatively small Reynolds number (Re < 400). Here, we note that the rings holding the laminarization screens penetrate into the flow channel, generating unsteady vortices. A screen step exists not only on the outer screen region, but also on the inner screen region. Using a new step-free screen, no critical transition is observed up to the highest flow rate achieved of Q = 103 lit/min. The original DMA widens the flow cross-section in the mixing region where the aerosol joins the sheath gas. The flow deceleration then arising at small aerosol input flow rates introduces another source of transfer function broadening, which, however, has negative resolution effects only at Q > 60 lit/min. This feature is suppressed here by modifying a single inlet piece. Although the two flow improvements implemented increase the resolving power in the analysis of very small particles, a substantial non-ideality of unclear origin remains: the best resolving power R found with electrosprayed ions of the protein Immunoglobulin is R = 13.9 for the trimer (IgG)3, and 12 for the monomer, even at a sheath/aerosol flow ratio of 100.

© 2017 American Association for Aerosol Research

Keywords:Jingkun Jiang
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号