首页 | 本学科首页   官方微博 | 高级检索  
     


Sequential injection titration of chloride in milk with potentiometric detection
Authors:M. J. Reis Lima, Sí  lvia M. V. Fernandes,Ant  nio O. S. S. Rangel
Affiliation:

a Escola Superior Agrária de Viseu, Instituto Superior Politécnico de Viseu, Campus Politécnico, Repeses, 3504-510, Viseu, Portugal

b Escola Superior de Biotecnologia, Universidade Católica Portuguesa, R. Dr. António Bernardino de Almeida, 4200-072, Porto, Portugal

Abstract:In this work, a sequential injection system for the pseudo-titration of chloride in milk is described. Milk was directly aspirated into the system and sandwiched between two silver nitrate plugs (titrant). The aspirated zones were then propelled to the detector (Ag2S/Ag tubular electrode), where the depletion in the titrant silver concentration (due to the formation of a AgCl precipitate) was monitored. The results obtained by the developed sequential injection titration method were not statistically different from those provided by the potentiometric reference procedure with relative standard deviations better than 3.4%. The sampling rates achieved were 17 samples per hour.
Keywords:Sequential injection   Milk   Chloride pseudo-titration
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏