首页 | 本学科首页   官方微博 | 高级检索  
     


Microscopical features in retrieved human Branemark implants: a report of 19 cases
Authors:A Piattelli  A Scarano  A Dalla Nora  G De Bona  GA Favero
Affiliation:Dental School, University of Chieti, Italy. apiattelli@unich.it
Abstract:The authors present a histologic analysis of 19 Branemark titanium implants retrieved for different causes: four implants were removed for abutment fracture, one for dental nerve dysesthesia, two for bone overheating, two for peri-implantitis, nine for mobility, one for unknown causes. In the implants removed for fracture a high bone-implant contact percentage was present (71.83 +/- 4.96%) with compact, mature bone at the interface. The picture of the failure due to bone overheating was characteristic with the presence of bone sequestra and of a gap between implant and bone filled by lymphocytes and plasma cells: many bacteria surrounded the necrotic bone and no newly regenerated bone was present. In peri-implantitis an inflammatory infiltrate was observed in the peri-implant tissues: a dense fibrous connective tissue was present around implants failed for mobility. The microscopical picture is certainly extremely important in identifying the causal determinants of an implant failure.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号