首页 | 本学科首页   官方微博 | 高级检索  
     


Ion recombination and polarity effect of ionization chambers in kilovoltage x-ray exposure measurements
Authors:IJ Das  SF Akber
Affiliation:Department of Radiation Oncology, Fox Chase Cancer Center, Philadelphia, Pennsylvania 19111, USA. i_das@fccc.edu
Abstract:Exposure measurements with ionization chambers are dependent on the correction factors related to the beam energy (ke), temperature and pressure (ktp), ionization recombination (Pion), and polarity (kpol) effects. In this work, six different chambers commonly used in diagnostic radiology were investigated for the Pion and kpol at various exposure rates by changing the tube voltage, beam current, exposure time, and distance. A special triaxial connector was used to connect chambers to an electrometer capable of measuring positive and negative polarity and 150 V and 300 V electrode potentials to measure the kpol and Pion, respectively. A mammography unit (24-35 kVp) and a diagnostic x-ray unit (60-125 kVp) were used. Results indicate that the magnitude of the Pion is linearly dependent on kVp for large volume (> 150 cm3) chambers and independent for small volume (< or = 150 cm3) chambers. In general, Pion is higher at higher exposures (increasing kVp, mAs, and decreasing distance); however, kpol is independent of exposure rate and kVp, but strongly depends on the sensitive volume of an ion chamber. Pion and kpol vary between 1-48% and 1-16%, respectively, among various chambers and exposure conditions. Chambers with larger volumes have higher values of Pion and kpol. The desired accuracy of +/- 5% in exposure measurements might not be feasible unless both the polarity and recombination effects are known and accounted accurately.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号