Development of a Position Sensitive Neutron Detector with High Efficiency and Energy Resolution for Use at High-Flux Beam Sources
Authors:Diane M Markoff  Vince Cianciolo  Chuck L Britton  Ronald G Cooper  Geoff L Greene
Affiliation:North Carolina State University, Department of Physics, Raleigh, NC;Oak Ridge National Laboratory, Oak Ridge, TN;Oak Ridge National Laboratory, Oak Ridge, TN;University of Tennessee, Knoxville, TN
Abstract:We are developing a high-efficiency neutron detector with 1 cm position resolution and coarse energy resolution for use at high-flux neutron source facilities currently proposed or under construction. The detector concept integrates a segmented 3He ionization chamber with the position sensitive, charged particle collection methods of a MicroMegas detector. Neutron absorption on the helium produces protons and tritons that ionize the fill gas. The charge is amplified in the field region around a wire mesh and subsequently detected in current mode by wire strips mounted on a substrate. One module consisting of a high-voltage plate, a field-shaping high-voltage plate, a grid and wire strips defines a detection region. For 100 % efficiency, detector modules are consecutively placed along the beam axis. Analysis over several regions with alternating wire strip orientation provides a two-dimensional beam profile. By using 3He, a 1/v absorption gas, each axial region captures neutrons of a different energy range, providing an energy-sensitive detection scheme especially useful at continuous beam sources.
Keywords:cold neutrons  ionization chamber  MicroMegas detector  neutron detector