首页 | 本学科首页   官方微博 | 高级检索  
     


Perchlorate-selective polymeric membrane electrode based on bis(dibenzoylmethanato)cobalt(II) complex as a neutral carrier
Authors:Rezaei Behzad  Meghdadi Soraia  Bagherpour Saeid
Affiliation:Department of Chemistry, Isfahan University of Technology, Isfahan 84156-83111, Islamic Republic of Iran. rezaei@cc.iut.ac.ir
Abstract:A synthesized bis(dibenzoylmethanato)Co(II) complex (Co(DBM)(2)), has been used as a ionophore for the preparation of a new perchlorate ion-selective electrode. The electrode exhibits a Nernstian response over the perchlorate concentration range of 8.0x10(-7)-1.0x10(-1)M with a slope of 60.3+/-0.5 mV per decade of concentration. The limit of detection as determined from the intersection of the extrapolated linear segments of the calibration plot is 5.6x10(-7)M. The electrode shows good selectivity towards perchlorate with respect to many common anions. The response time of the sensor is very fast (< or = 5s), and can be used for at least 2 months in the pH range of 2.0-9.0. The electrode was used to determine perchlorate in water and human urine. The interaction of the ionophore with perchlorate ions was demonstrated by UV-vis spectroscopy.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!