首页 | 本学科首页   官方微博 | 高级检索  
     

岛津GC-14C气相色谱仪故障分析
引用本文:周承强,焦志峰,刘克萍,徐红.岛津GC-14C气相色谱仪故障分析[J].广东化工,2012,39(10):84-85.
作者姓名:周承强  焦志峰  刘克萍  徐红
作者单位:克拉玛依石化公司,新疆,834003
摘    要:文章较详细地介绍了测定酸酐反应气组成岛津GC-14c气相色谱仪基线电压频繁出现突然掉落至-5000ΜV的故障现象、故障分析、检修方法及修理过程中的注意事项。同时解决了顺酐反应气组分含量测定不准的难题。

关 键 词:气相色谱仪  岛津  基线  重复性

Shimadzu GC-14 C Gas Chromatograph Failure Analysis
Zhou Chengqiang,Jiao Zhifeng,Liu Keping,Xu Hong.Shimadzu GC-14 C Gas Chromatograph Failure Analysis[J].Guangdong Chemical Industry,2012,39(10):84-85.
Authors:Zhou Chengqiang  Jiao Zhifeng  Liu Keping  Xu Hong
Affiliation:(Kelamayi Petrochemical Company, Xinjiang 834003, China)
Abstract:The paper introduced in detail the determination of maleic anhydride reaction gas i Shimadzu GC-14 C gas chromatograph baseline voltage frequent suddenly dropped into -5000 MV malfunctions maintenance methods of fault analysis and repair the matters of attention in the process, it solves the maleic anhydride reaction gas content determination of the problem to components.
Keywords:gas chromatography  Shimadzu  baseline  repeatability