Triple-path collector optics for grazing incident x-ray emission spectrometer |
| |
Authors: | Tokushima T Horikawa Y Shin S |
| |
Affiliation: | RIKEN/SPring-8, 1-1-1 Kouto, Sayo, Hyogo, 679-5148, Japan. |
| |
Abstract: | A new type of collector optics was developed for grazing incident x-ray emission spectrometer. The collector optics used two cylindrical mirrors to add two extra light paths while keeping the center light path that directly illuminates the grating. The design and properties of the spectrometer using the triple-path collector optics were evaluated using ray-tracing simulations, and validity of this design in terms of throughput and energy resolution was confirmed by the experimentally obtained spectra. |
| |