首页 | 本学科首页   官方微博 | 高级检索  
     


A gated liquid-scintillator-based neutron detector for fast-ignitor experiments and down-scattered neutron measurements
Authors:Stoeckl C  Cruz M  Glebov V Yu  Knauer J P  Lauck R  Marshall K  Mileham C  Sangster T C  Theobald W
Affiliation:Laboratory for Laser Energetics, University of Rochester, 250 East River Road, Rochester, New York 14623, USA. csto@lle.rochester.edu
Abstract:The detection of neutrons in fast-ignitor experiments or down-scattered neutrons in inertial fusion experiments is very challenging since it requires the neutron detection system to recover within 10-100 ns from a high background orders of magnitude stronger than the signal of interest. The background is either the hard x-ray emission from short-pulse laser target interactions for the fast-ignitor experiments or the primary neutron signal for the down-scattered neutrons. A liquid-scintillator detector has been developed using a gated photomultiplier that suppresses the background signal and eliminates the afterglow present in conventional plastic scintillators.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号