Determination of appropriate RF blocking impedance for MRI surface coils and arrays |
| |
Authors: | Armen Kocharian Phillip J. Rossman Thomas C. Hulshizer Joel P. Felmlee Stephen J. Riederer |
| |
Affiliation: | (1) Department of Diagnostic Radiology, Mayo Clinic and Foundation, MR Research Laboratory, 200 First St SW, 55905 Rochester, MN, USA |
| |
Abstract: | Surface and phased array receiving coils in MRI typically require that RF excitation be accomplished using the body coil. This process requires that the receiving coils contain blocking circuitry to increase the overall circuit impedance during RF excitation and withstand the electromotive force induced by the applied electromagnetic field. The aim of this study was to determine the optimal impedance range required during RF excitation based on an assessment of image quality. The experimental results are fit by an exponential model and establish criteria that can be applied for general receiver coil design. |
| |
Keywords: | MRI Surface coil Blocking circuit |
本文献已被 PubMed SpringerLink 等数据库收录! |