首页 | 本学科首页   官方微博 | 高级检索  
     


Corrosion susceptibility of implant materials Ti-5Al-4V and Ti-6Al-4Fe in artificial extra-cellular fluids
Authors:Mihai V Popa  Ecaterina Vasilescu  Paula Drob  Julia Mirza-Rosca  Cora Vasilescu
Affiliation:a Institute of Physical Chemistry “I.G. Murgulescu”, Spl. Independentei 202, P.O. Box 12-194, 060041 Bucharest, Romania
b Politehnica University, Faculty of Chemistry, Str. Polizu Nr.1, 78126 Bucharest, Romania
c Las Palmas de Gran Canaria University, Despacho 11, 35017 Tafira, Spain
Abstract:The paper is aimed at detecting the corrosion susceptibility (generated by the potential gradients, due to the pH changes) of implant materials Ti-5Al-4V and Ti-6Al-4Fe alloys exposed in extra-cellular fluids for long term (10 000 exposure hours) in comparison with pure titanium. The titanium and its ternary alloys exhibited spontaneous passivity in Ringer’s solutions of pH=6.98, 4.35 and 2.5, simulating the real situations that can arise in surgical applications. Potential gradients determined in presumptive extreme pH conditions indicate that there is no probability for local corrosion. Impedance spectra were fitted with one time constant equivalent circuit, typically for protective, compact oxide film. The fitting parameters indicate long-term stability of the passive layers in surgical implant conditions. Atomic force microscopy (AFM) and spectral infrared (IR) data are important arguments in supporting the conclusion that titanium and its ternary alloys have a very good corrosion resistance in long-term functional conditions.
Keywords:Surface properties   Thin films   Titanium alloys
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏