An image reconstruction algorithm of EIT based on pulmonary prior information |
| |
Authors: | Huaxiang Wang Li Hu Jing Wang Lu Li |
| |
Affiliation: | (1) School of Electrical Engineering and Automation, Tianjin University, Tianjin, 300072, China |
| |
Abstract: | Using a CT scan of the pulmonary tissue, a human pulmonary model is established combined with the structure property of the human lung tissue using the software COMSOL. Combined with the conductivity contribution information of the human tissue and organ, an image reconstruction method of electrical impedance tomography based on pulmonary prior information is proposed using the conjugate gradient method. Simulation results show that the uniformity index of sensitivity distribution of the pulmonary model is 15.568, which is significantly reduced compared with 34.218 based on the round field. The proposed algorithm improves the uniformity of the sensing field, the image resolution of the conductivity distribution of pulmonary tissue and the quality of the reconstruction image based on pulmonary prior information. __________ Translated from Journal of Tianjin University, 2008, 41(4): 383–388 [译自: 天津大学学报] |
| |
Keywords: | electrical impedance tomography (EIT) prior information pulmonary model of human image reconstruction COMSOL |
本文献已被 万方数据 SpringerLink 等数据库收录! |
|