首页 | 本学科首页   官方微博 | 高级检索  
     


Contactless technique for the measurement of electrical resistivity in anisotropic materials
Authors:Zeller C  Denenstein A  Foley G M
Affiliation:Moore School of Electrical Engineering and Laboratory for Research on the Structure of Matter, University of Pennsylvania, Philadelphia, Pennsylvania 19104.
Abstract:Measurement of basal plane electrical resistivities in quasi-two-dimensional materials by conventional four-point bridge techniques is extremely difficult. In making measurements of room temperature basal plane resistivities in highly conducting highly anisotropic synthetic metals we have developed an rf inductive technique which is both simple to use and provides accurate results. At the frequency we employ (100 kHz) the method is appropriate for resistivities in the range 10 (-3)-10(-7)Omega cm. Samples are thin square plates 5 mm on a side. The system is calibrated by fitting data from a series of samples of known resistivity to semi-empirical formulae. The theoretical motivation leading to these expressions is discussed and extensive calibration data are presented. The advantages of rf techniques in general have already been discussed. Two advantages commend the particular technique we describe here. First, it is simple in design and easily constructed. Second, with careful calibration, highly accurate resistivity measurements can be made in a very straightforward manner.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号