首页 | 本学科首页   官方微博 | 高级检索  
     


AlGaAs DH pump laser for photoluminescence lifetime measurements
Authors:Nelson R J
Affiliation:Bell Laboratories, Murray Hill, New Jersey 07974.
Abstract:The use of GaAlAs double heterostructure lasers as a pulsed excitation source for photoluminescence time-decay measurements is described. Subnanosecond laser pulses easily allow the determination of luminescence decay times >/=500 ps using a single photon counting system. In contrast to mode-locked gas or dye lasers, this new technique utilizes simple equipment (diode laser and pulse generator) and requires no special alignment or tuning procedures.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号