首页 | 本学科首页   官方微博 | 高级检索  
     


Thin-film electrode based droplet detection for microfluidic systems
Authors:EV MoiseevaAuthor VitaeAA FletcherAuthor Vitae  CK HarnettAuthor Vitae
Affiliation:University of Louisville, Department of Electrical and Computer Engineering, Louisville, KY 40208, USA
Abstract:We report on a droplet-producing microfluidic system with electrical impedance-based detection. The microfluidic devices are made of polydimethylsiloxane (PDMS) and glass with thin film electrodes connected to an impedance-monitoring circuit. Immiscible fluids containing the hydrophobic and hydrophilic phases are injected with syringe pumps and spontaneously break into water-in-oil droplet trains. When a droplet passes between a pair of electrodes in a medium having different electrical conductivity, the resulting impedance change signals the presence of the particle for closed-loop feedback during processing. The circuit produces a digital pulse for input into a computer control system. The droplet detector allows estimation of a droplet's arrival time at the microfluidic chip outlet for dispensing applications. Droplet detection is required in applications that count, sort, and direct microfluidic droplets. Because of their low cost and simplicity, microelectrode-based droplet detection techniques should find applications in digital microfluidics and in three-dimensional printing technology for rapid prototyping and biotechnology.
Keywords:Droplet microfluidics  PDMS  Lab on chip  Droplet detection circuit
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏