首页 | 本学科首页   官方微博 | 高级检索  
     


Thermal diffusivity measurement of CVD diamond film using a step heating technique
Authors:Y. Li  R. E. Taylor  A. Nabi
Affiliation:(1) Thermophysical Properties Research Laboratory, School of Mechanical Engineering, Purdue University, 47907 West Lafayette, Indiana, USA;(2) Present address: Department of Agricultural Engineering, Purdue University, 47907 West Lafayette, Indiana, USA;(3) Aeronautical Research Center, Technion-Israel Institute of Technology, 32000 Haifa, Israel
Abstract:A step heating method for the measurement of the thermal diffusivity of diamond thin film is described. The step heating method is a transient heat flow method. Transient temperature profiles are generated in a strip-shaped sample by heating one end of the sample while the other end is clamped to a heat sink. Three thermocouples are used along the heat path. The results are compared with the literature values over the temperature range from –190 to 50°C.
Keywords:diamond film  low temperatures  step heating technique  thermal conductivity  thermal diffusivity
本文献已被 SpringerLink 等数据库收录!