Design of efficient, broadband single-element (20-80 MHz) ultrasonic transducers for medical imaging applications |
| |
Authors: | Cannata Jonathan M Ritter Timothy A Chen Wo-Hsing Silverman Ronald H Shung K Kirk |
| |
Affiliation: | Dept. of Biomed. Eng., Univ. of Southern California, Los Angeles, CA, USA; |
| |
Abstract: | This paper discusses the design, fabrication, and testing of sensitive broadband lithium niobate (LiNbO/sub 3/) single-element ultrasonic transducers in the 20-80 MHz frequency range. Transducers of varying dimensions were built for an f# range of 2.0-3.1. The desired focal depths were achieved by either casting an acoustic lens on the transducer face or press-focusing the piezoelectric into a spherical curvature. For designs that required electrical impedance matching, a low impedance transmission line coaxial cable was used. All transducers were tested in a pulse-echo arrangement, whereby the center frequency, bandwidth, insertion loss, and focal depth were measured. Several transducers were fabricated with center frequencies in the 20-80 MHz range with the measured -6 dB bandwidths and two-way insertion loss values ranging from 57 to 74% and 9.6 to 21.3 dB, respectively. Both transducer focusing techniques proved successful in producing highly sensitive, high-frequency, single-element, ultrasonic-imaging transducers. In vivo and in vitro ultrasonic backscatter microscope (UBM) images of human eyes were obtained with the 50 MHz transducers. The high sensitivity of these devices could possibly allow for an increase in depth of penetration, higher image signal-to-noise ratio (SNR), and improved image contrast at high frequencies when compared to previously reported results. |
| |