首页 | 本学科首页   官方微博 | 高级检索  
     


Protein sizing on a microchip
Authors:Bousse L  Mouradian S  Minalla A  Yee H  Williams K  Dubrow R
Affiliation:Caliper Technologies Corp., Mountain View, California 94043-2234, USA. luc.bousse@calipertech.com
Abstract:We have developed a microfabricated analytical device on a glass chip that performs a protein sizing assay, by integrating the required separation, staining, virtual destaining, and detection steps. To obtain a universal noncovalent fluorescent labeling method, we have combined on-chip dye staining with a novel electrophoretic dilution step. Denatured protein-sodium dodecyl sulfate (SDS) complexes are loaded on a chip and bind a fluorescent dye as the separation begins. At the end of the separation channel, an intersection is used to dilute the SDS below its critical micelle concentration before the detection point. This strongly reduces the background due to dye molecules bound to SDS micelles and also increases the peak amplitude by 1 order of magnitude. Both the on-chip staining and SDS dilution steps occur in the 100-ms time scale and are approximately 10(4) times faster than their conventional counterparts in SDS-PAGE. This represents a much greater speed increase due to microfabrication than has been obtained in other assay steps such as electrophoretic separations. We have designed and tested a microchip capable of sequentially analyzing 11 different samples, with sizing accuracy better than 5% and high sensitivity (30 nM for carbonic anhydrase).
Keywords:
本文献已被 PubMed 等数据库收录!