首页 | 本学科首页   官方微博 | 高级检索  
     

SVM3000斯塔宾格黏度计测定液体石油产品的运动黏度
引用本文:宋昌盛,王群威,王晶,徐臻瑾,满国瑜,谈啸.SVM3000斯塔宾格黏度计测定液体石油产品的运动黏度[J].分析仪器,2013(6):18-21.
作者姓名:宋昌盛  王群威  王晶  徐臻瑾  满国瑜  谈啸
作者单位:[1]太仓出入境检验检疫局,太仓215400 [2]宁波出入境检验检疫局,宁波315012 [3]重庆出入境检验检疫局,重庆400020 [4]中国石油华东润滑油厂,太仓215433 [5]奥地利安东帕中国有限公司,上海,200233
基金项目:国家认证认可监督管理委员会标准制(修)订项目(2009B616)。
摘    要:利用奥地利Anton Paar公司生产的SVM 3000斯塔宾格黏度计(Stabinger viscometer)测定了基础油、润滑油、燃料油和原油等透明和不透明石油产品的运动黏度。对不同黏度等级基础油的重复测定结果符合ASTM D7042规定的重复性限要求,三个实验室对润滑油、燃料油和原油的测定结果相对标准偏差在1%以下。分别用斯塔宾格黏度计和毛细管黏度计测定了多种基础油的运动黏度,结果表明两种方法测定40℃运动黏度的结果相关性为Y=0.9995 X (Y 为斯塔宾格黏度计的测定结果,X为毛细管黏度计的测定结果),100℃运动黏度的结果相关性为Y=1.0033 X ,说明两种方法的测定结果一致,无显著性差异。

关 键 词:SVM3000  斯塔宾格黏度计  石油产品  毛细管黏度计

Determination of kinematic viscosity of petroleum products by SVM 3000 Stabinger viscometer
Song Changsheng,Wang Qunwei,Wang Jing,Xu Zhenjin,Man Guoyu,Tan Xiao.Determination of kinematic viscosity of petroleum products by SVM 3000 Stabinger viscometer[J].Analytical Instrumentation,2013(6):18-21.
Authors:Song Changsheng  Wang Qunwei  Wang Jing  Xu Zhenjin  Man Guoyu  Tan Xiao
Affiliation:1. Taicang Entry-Exit Inspection and Quarantine Bureau, Taicang 215400,China;2. Ningbo Entry-Exit Inspection and Quarantine Bureau, Ningbo 315012, China; 3. Chongqing Entry-Exit Inspection and Quarantine Bureau, Chongqing 400020,China ;4. Petro China Huadong Lubricant Blending Plant, Taicang 215433, China; 5. Anton Paar China, Shanghai 200233, China)
Abstract:The SVM 3000 Stabinger viscometer provided by Anton Paar GmbH ,is used to determine the kinematic viscosity of petroleum products such as base oil ,lubricant oil ,fuel oil and crude oil. The successive testing results of several types of base oil with different viscosity grade ,comply with the repeat-ability limits of ASTM D7042. Furthermore ,the RSD values for lubricant oil ,fuel oil and crude oil from three different laboratories are all below 1%. In addition ,the kinematic viscosity of base oils is determined by Stabinger viscometer and glass capillary viscometer respectively. The correlation analysis between two methods show that Y=0.9995 X at 40℃ ,and Y=1.0033 X at 100℃ ,w herein X and Y are the testing re-sults by glass capillary viscometer and stabinger viscometer respectively. It suggests there is no obvious bi-as between two methods.
Keywords:SVM 3000  Stabinger viscometer  petroleum product  glass capillary viscometer
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号