首页 | 本学科首页   官方微博 | 高级检索  
     


Spiral scanning with anisotropic field of view
Authors:KF King
Affiliation:GE Medical Systems, Milwaukee, Wisconsin 53201, USA.
Abstract:Spiral scanning has been used to achieve much shorter scan times than conventional techniques for a wide range of applications. The major drawback with spiral scans is blurring from off-resonant spins, which is proportional to the readout time. Blurring limits maximal spatial resolution and minimal scan time potentially achievable with spiral scanning. Anisotropic field of view is used in conventional scanning to improve image quality by matching k-space trajectory to object characteristics. Anisotropic field of view improves spatial resolution in spiral scanning without increasing scan time or blurring. The resolution improvement results from increased maximal k-space radius allowed by the lower field of view. A field of view reduction by a factor of 2 in one direction provides up to 60% resolution improvement in that direction. Reduced SNR also results from non-uniform k-space sampling.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号