Method for measuring the losses and loading of a quartz crystal microbalance |
| |
Authors: | Kankare Jouko Loikas Kari Salomäki Mikko |
| |
Affiliation: | Department of Chemistry, University of Turku, FIN-20014 Turku, Finland. jouko.kankare@utu.fi |
| |
Abstract: | A novel, experimentally simple, and highly sensitive method for measuring the loading of a quartz crystal resonator was developed. The method is based on the use of double-sideband suppressed-carrier modulated high-frequency signal, which is swept through the resonance range of the resonator. Induced current in the resonator is passed through a capacitor, and the voltage over the capacitor is demodulated on an analog multiplier. The phase and amplitude information is carried to the frequency-doubled modulation signal and measured on a conventional low-frequency two-phase lock-in amplifier. A complex dimensionless loading parameter is obtained from the experimental data by nonlinear model fitting. The real and complex parts of this loading parameter have a simple relationship with other parameters commonly used for characterizing the resonator loading. The performance of the method was demonstrated by measuring a series of different glycerol-water mixtures ranging from 0 up to 100% glycerol. The results were close to the shear acoustic impedance of these mixtures measured and calculated from their viscosities and densities. |
| |