Biosensor for arsenite using arsenite oxidase and multiwalled carbon nanotube modified electrodes
Authors:Male Keith B  Hrapovic Sabahudin  Santini Joanne M  Luong John H T
Affiliation:Biotechnology Research Institute, National Research Council Canada, Montreal, Quebec, Canada H4P 2R2.
Abstract:A biosensor for arsenite has been developed using molybdenum-containing arsenite oxidase, prepared from the chemolithoautotroph NT-26 that oxidizes arsenite to arsenate. The enzyme was galvanostatically deposited for 10 min at 10 microA onto the active surface of a multiwalled carbon nanotube modified glassy carbon electrode. The resulting biosensor enabled direct electron transfer, i.e., effecting reduction and then reoxidization of the enzyme without an artificial electron-transfer mediator. Arsenite was detected within 10 s at an applied potential of 0.3 V with linearity up to 500 ppb and a detection limit of 1 ppb. The biosensor exhibited excellent reproducibility, 2% at 95% confidence interval for 12 repeated analyses of 25 ppb arsenite. Copper, a severe interfering species commonly found in groundwater, did not interfere, and the biosensor was applicable for repeated analysis of spiked arsenite in tap water, river water, and a commercial mineral water.