首页 | 本学科首页   官方微博 | 高级检索  
     

医院中子照射器分析室剂量场研究
引用本文:张紫竹,宋明哲,李玮,陈军,杨勇,李义国. 医院中子照射器分析室剂量场研究[J]. 原子能科学技术, 2012, 46(Z1): 623-625. DOI: 10.7538/yzk.2012.46.suppl.0623
作者姓名:张紫竹  宋明哲  李玮  陈军  杨勇  李义国
作者单位:1.中国核工业北京四〇一医院,北京102413;2.中国原子能科学研究院 放射性计量测试部,北京102413;3.中国原子能科学研究院 反应堆工程设计研究所,北京102413;4.北京凯佰特科技有限公司,北京 102413
摘    要:医院中子照射器建成后,对分析室内及其屏蔽门外的γ剂量率和中子剂量当量率进行了测量,测量结果显示:分析室内局部γ剂量率与设计值相差较大,分析室屏蔽门外γ剂量率超过原设计监督区限值7.5 μSv/h,因此需对分析室内部及其屏蔽门进行屏蔽改造。根据蒙特卡罗程序模拟计算结果及实际使用情况给出最终屏蔽方案,即在分析室束流孔道所在墙面加装厚度为16 cm的铅屏蔽材料屏蔽γ射线,对四周墙面及屏蔽门内侧加装厚度为1 cm的含锂聚乙烯板屏蔽散射中子。改造后分析室剂量最高点γ剂量率下降277倍,中子剂量当量率下降5.8倍,屏蔽门外γ剂量率下降近90倍。

关 键 词:医院中子照射器   &gamma   剂量率   中子剂量当量率   屏蔽改造

Dose Field Research of Analysis Room for In-Hospital Neutron Irradiator
ZHANG Zi-zhu , SONG Ming-zhe , LI Wei , CHEN Jun , YANG Yong , LI Yi-guo. Dose Field Research of Analysis Room for In-Hospital Neutron Irradiator[J]. Atomic Energy Science and Technology, 2012, 46(Z1): 623-625. DOI: 10.7538/yzk.2012.46.suppl.0623
Authors:ZHANG Zi-zhu    SONG Ming-zhe    LI Wei    CHEN Jun    YANG Yong    LI Yi-guo
Affiliation:1.China Nuclear Industry Beijing 401 Hospital, Beijing 102413, China;2.China Institute of Atomic Energy, P. O. Box 275-20, Beijing 102413, China;3.China Institute of Atomic Energy, P. O. Box 275-75, Beijing 102413,China;4.Beijing Capture Technology Co., Ltd., Beijing 102413, China
Abstract:Neutron equivalent dose rate and γ ray dose rate inside the analysis room of the in hospital neutron irradiator (IHNI) and outdoor were measured. The results show that γ ray dose rate inside the analysis room exceeds calculation value many times and γ ray dose rate outdoor is higher than supervision region dose limit of 7.5 μSv/h. According to the measurement results and the Monte Carlo simulation, the following shielding plan was adopted. Lead shielding with thinkness of 16 cm was installed on the wall, which faces the neutron beam, to shield γ ray, and lithium polyethylene plate with thinkness of 1 cm was installed on all the wall (not including ceiling and floor) to shield scattering neutron. After shielding transformation, the highest γ ray dose rate point inside the analysis room decreased 277 times, the neutron equivalent dose rate decreased 5.8 times, and the outdoor γ ray dose rate decreased nearly 90 times.
Keywords:in-hospital neutron irradiator  γray dose rate  neutron equivalent dose rate  shielding transformation
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《原子能科学技术》浏览原始摘要信息
点击此处可从《原子能科学技术》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号