首页 | 本学科首页   官方微博 | 高级检索  
     


X-ray imaging using amorphous selenium: determination of Swank factor by pulse height spectroscopy
Authors:IM Blevis  DC Hunt  JA Rowlands
Affiliation:Sunnybrook Health Science Centre, University of Toronto, Ontario, Canada. ira@srcl.sunnybrook.utoronto.ca
Abstract:The use of photoconductors, especially amorphous selenium (a-Se), in x-ray imaging is currently of interest. A critical performance parameter of an imaging detector is the Swank factor for degradation of the signal to noise ratio, or DQE(0), due to variations in the detector response. The Swank factor is evaluated from measured pulse height spectra generated by the absorption of monoenergetic x-ray photons. The spectra show an additional width over previous theoretical expectations, but the Swank factor is still close to the high values previously predicted theoretically.
Keywords:
本文献已被 PubMed 等数据库收录!