首页 | 本学科首页   官方微博 | 高级检索  
     


Wear Performance of Laser Processed Tantalum Coatings
Authors:Dittrick Stanley  Balla Vamsi Krishna  Bose Susmita  Bandyopadhyay Amit
Affiliation:W. M. Keck Biomedical Materials Research Laboratory, School of Mechanical and Materials Engineering, Washington State University, Pullman, WA 99164-2920, USA.
Abstract:This first generation investigation evaluates the in vitro tribological performance of laser-processed Ta coatings on Ti for load-bearing implant applications. Linear reciprocating wear tests in simulated body fluid showed one order of magnitude less wear rate, of the order of 10(-4)mm(3)(N.m)(-1), for Ta coatings compared to Ti. Our results demonstrate that Ta coatings can potentially minimize the early-stage bone-implant interface micro-motion induced wear debris generation due to their excellent bioactivity comparable to that of hydroxyapatite (HA), high wear resistance and toughness compared to popular HA coatings.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!