An ultralow-power switched opamp-based 10-B integrated ADC for implantable biomedical applications
Authors:Bonfini  G Brogna  AS Garbossa  C Colombini  L Bacci  M Chicca  S Bigongiari  F Guerrini  NC Ferri  G
Affiliation:Dipt. di Ingegneria dell'Informazione, Univ. degli Studi di Pisa, Italy;
Abstract:This paper describes an ultralow-power switched opamp-based integrated analog-to-digital converter (ADC) for cardiac pacemakers applications. The ADC consumption, measured on 10 chip samples and averaged, is 8.18 /spl mu/W (stand-by value: 1 nW) for the analog part and of 9.71 /spl mu/W (5 nW) for the digital one, using a supply battery of 2.8 V. The converter has a resolution of 10-b, its typical operating clock frequency is 32 KHz (2.9 KS/s sampling rate) and is able to reach the same resolution at 2 V (0.7 KS/s sampling rate), with a dissipation of 1 /spl mu/W and 1.3 /spl mu/W for analog and digital part, respectively.