Influence of temperature on the emitted dose of an oral metered dose inhaler
Authors:RO Williams  MK Barron
Affiliation:University of Texas at Austin, College of Pharmacy, Pharmaceutics Division 78712-1074, USA. williro@mail.utexas.edu
Abstract:The performance of metered dose inhalers is critical for the efficient delivery of drugs to the intended site of deposition in the respiratory tract. The temperature at which metered dose inhaler products are used by patients may influence the physicochemical characteristics of the emitted dose. Product performance characteristics of a metered dose inhaler containing beclomethasone dipropionate and oleic acid in a blend of chlorofluorocarbon propellants, Freon-11 and Freon-12, were determined by cascade impaction analysis and dose delivery through the valve after the metering chamber was loaded and actuated at 4 degrees C, 23 degrees C, and 40 degrees C. The dose delivered from the valve was not affected by the temperature at which the metering chamber was loaded and actuated. The mass median aerodynamic particle size of the emitted aerosol decreased and the percentage respirable fraction increased as the temperature was increased. The geometric standard deviation of the particle size distribution was not significantly affected by the temperature at which the metering chamber was loaded and actuated. The temperature at which a metered dose inhaler is used by a patient may influence the amount of drug that is potentially respirable; therefore, the dose expected to be delivered and the corresponding therapeutic effect may also be affected.