首页 | 本学科首页   官方微博 | 高级检索  
     

等离子喷涂硼化镧阴极电子枪
引用本文:王厚稳,朱以漳.等离子喷涂硼化镧阴极电子枪[J].原子能科学技术,1989,23(4):52-56.
作者姓名:王厚稳  朱以漳
作者单位:南京大学 (王厚稳,朱以漳),南京大学(陈树群)
摘    要:一、引言现代大型医用电子直线加速器电子枪的设计有如下特点: 1.要求电子枪束流在大范围内连续可调。对于一个二极型电子枪系统,枪必须工作在温度限制区域内,故阴极发射的稳定性十分重要。2.行波电子直线加速器通常采用可拆性结构,这就要求阴极必须有较好的抗中毒特性。

关 键 词:电子枪  等离子喷涂  硼化镧  阴极

AN ELECTRON GUN WITH A PLASMA SPRAY LaB_6 CATHODE
WANG HOUWEN,ZHU YIZHANG,CHEN SHUQUN Nanjing University.AN ELECTRON GUN WITH A PLASMA SPRAY LaB_6 CATHODE[J].Atomic Energy Science and Technology,1989,23(4):52-56.
Authors:WANG HOUWEN  ZHU YIZHANG  CHEN SHUQUN Nanjing University
Affiliation:WANG HOUWEN;ZHU YIZHANG;CHEN SHUQUN Nanjing University
Abstract:The structure and characteristic of an electron gun with a plasma spray LaB_6cathode for medical electron linear accelerator are described in the paper. Theemission, poison and stability performances of scandium salt Ba-W, plasma sprayLaB_6, polycrystalline LaB_6 and pure Ta cathodes are measured. The results showthat the plasma spray LaB_6 cathode is more suitable for the medical electron linearaccelerator with detachable vacuum system than others.
Keywords:Medical electron linear accelerator  Electron gun  Plasma spray LaB_6 cathode  Anti-poison performance
本文献已被 CNKI 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号