Effect of the Response Delay of the Measuring System on Thermal Diffusivity Measurement Using the Flash Method1 |
| |
Authors: | N. Araki D. W. Tang M. Suzuki A. Makino |
| |
Affiliation: | (1) Department of Mechanical Engineering, Faculty of Engineering, Shizuoka University, Hamamatsu, 432-8561, Japan |
| |
Abstract: | This work investigates the effect of the response delay of a measuring system on a thermal diffusivity measurement. A model of an m th-order delay in the measuring system is introduced, and a general expression for the output of the system with temperature response as input is derived. The effect on the temperature response caused by such a system is discussed. As a practical example, a third-order measuring system is considered. The measured temperature responses of stainless steel foils are compared with those calculated with the model of a third-order delay system. Good agreement between the two results is shown. |
| |
Keywords: | flash method response delay thermal diffusivity time constant |